FAST 6000™

 

FAST 6000, Rapid portable fluorometric assay system


Rapid, Portable, Fluorometric
Assay System

The Flow Assay and Sensing System (FAST) is a second-generation instrument, developed under contract to the Naval Research Laboratory (NRL) as a rapid, portable system for performing flow immunoassays with detection limits to 1 part per billion. The FAST 6000 can be configured either as a single channel, single analyte instrument or as a six channel, six analyte instrument, capable of performing six simultaneous assays for six different analytes on the same sample. The system is completely self-contained, needing only a bottle of buffer and a waste bottle for field operation.

In operation, this small-molecule assay system uses a fluorophore-labeled analyte, bound to analyte-specific antibody that is immobilized on a permeable membrane. When a sample containing analyte passes through the membrane an exchange reaction occurs, displacing the labeled analyte which is measured downstream. A typical assay requires two minutes to run and up to fifty assays can be performed before the instrument must be recharged.

FAST 6000™ System Specifications

Characteristic

Description

Assay Medium

Membrane-bound antibody.

Assay Vessel

Disposable coupon.

Limit of Detection

1 to 5 ppb typical.

Minimum Assay Time

120 sec

Sample Size

750 µl min for 6-channel; 150 µl min for 1-channel.

Operational Flow Rate

0 to 1.5 ml/min; user programmable.

Optical Source

5 mW solid-state laser, 635 nm.

Optical Detection

Displacement fluorescence assay.

Photocurrent Resolution

0.076 pA

Maximum Photocurrent

2,500 pA

Measurement Rate

One reading per sec, or slower; controlled by user.

Warm-Up Time

15 min

Number of Channels

One or six (specified at time of order).

PC Interface Requirements

Windows® 95 compatible computer.

Hardware Requirements

Same as those for Windows® 95.

Power Requirements

6.9 W (386, air pump, and liquid pumps on), supplied by rechargeable internal Li-ion battery pack or from external lump-in-cord wall supply.

Size

24.1 cm L x 7.6 cm H x 27.3 cm W

Weight

3.68 kg

Research International reserves the right to change specifications on any of the devices
mentioned without prior notification.

For more information: